Recommendations for dilution of G135 developer concentrate for film processing (according to the icons located on the cans).
The G135 developer retail package (1 box) consists of two identical semi-kits, each of which, in turn, consists of three components. Each of the semi-sets can be diluted up to 20 liters of working solution. To dilute one half of the developer, it is necessary to pour 5 liters of concentrate (component A) into a container with 10 liters of water (water temperature from + 15 ° C to + 40 ° C), add 0.5 liters of concentrate (component B) and add 0.5 liters of concentrate (component C). After that, add the remaining 4 liters of water to the resulting solution.
ATTENTION! After each addition of the next component of the concentrate, the solutions should be thoroughly mixed.
Storage conditions for AGFA NDT chemicals:
During storage, chemical reagents should have optimal temperature limits, which, in accordance with regulatory and technical documentation, constitute a temperature range from + 4ºС to + 23ºС.
The positive temperature range of application of these chemicals is primarily associated with a high water content in their composition. These chemical reagents at low temperatures (up to -50 ° C) slightly change their physico-chemical properties.
In case of freezing of chemical reagents, it is necessary to thaw them at room temperature and use them as usual. The same applies to chemical reagents of the STRUCTURIX brand for automatic X-ray film processing.
The table shows data on the configuration and consumption of chemical reagents. These data will also allow to determine the total weight and dimensions when delivering the required amount of chemicals.